Oakton Temp 9500, Advanced, PID, Benchtop Temperature Controllers, Oakton, Instruments

Oakton Temp 9500 Advanced PID Benchtop Temperature Controllers

  • Oakton Temp 9500, Advanced, PID, Benchtop Temperature Controllers, Oakton, InstrumentsOutlet receptacle allows for direct plug-in of heating devices such as hot plates, heat tapes, or mantles.

  • Features USB output and software with real-time graphing and simple setup of ramp and soak programming.

These temperature controllers can be used for a wide variety of applications including pilot process plants, R & D labs, or for OEM requirements. Rear panel output receptacle for direct plug-in of heaters and other resistive devices. Field calibrations help to improve system accuracy by entering the offset value to correct for individual probe error.

Simultaneously view both the measured and set point on the two-line alphanumeric display.Temperature scale selectable to read in F, C, K (Kelvin), or R (Rankine). Front panel LEDs indicate output and alarm conditions.

Advanced models feature heater output capacity of 1750/3450 watts max, ramp-and-soak profiling for more complicated processes, USB computer interface, free software that features real-time graphing and simple setup of ramp and soak progamming, alarm relay with adjusted hysteresis, and 4 to 20 mA and 1 to 5 V outputs for connection to a recorder or data logger. In addition to thermocouple inputs, advanced models also accept RTD and thermistor inputs.

The 115 VAC models include a 6-ft (1.8-m) power cord and US standard plug and receptacle; the 230 VAC models feature an IEC cord set and universal receptacle. Built-in grid support bracket allows the unit to be mounted onto a laboratory frame or grid.


Safety featuresOutput power to load device is automatically shut down in the event of a broken sensor or control loop break. Overtemperature protection shuts down the system if the user-settable over-temperature or timer is exceeded. Outlet power receptacle is fused separately from the controller for added safety. Audible and visual out-of-range alarms are also included.

Control type On/Off, Programmable PID, autotune PID
Ramp/Soak Yes
Temperature Sensor Thermocouple types J,K,T,E,N,B,R,S Thermistor with 1/4" phono plugs RTDs with 3-pin or Oakton/Digi-Sense connectors
OutputControl 4 to 20/20 to 4 selectable, 1 to 5 VDC
OutputCurrent 15 amps maximum

Control type: on/off, programmable PID, autotune PID
Resolution: 0.1; 1 above 999.9 and below 99.9
Accuracy: Types J, K, T, E, and N: 0.1% of reading, 0.7F (0.4C) above 248F (100C); 0.1% of reading, 2F (1C) below 248F (100C) Types B, R, and S: 0.1% of reading, 1.8F (1.0C) Thermistors and RTDs: 0.1% of reading, 0.7F (0.4C)
Display: LCD
Digital output: USB
Operating ambient: 32 to 104F (0 to 40C); 0 to 90% RH, noncondensing
Control output: powered receptacle rated for 115/230 VAC, 15 A max
Alarm output: one SPDT relay rated for 230 VAC, 2 A max; resistive
Dimensions (W x H x D): 8" x 33/4" x 9" (20.3 x 9.5 x 22.9 cm)

Back to Temperature Controllers